The Testomat® MODUL Monochloramin (NH2CL) is an online
analyzer that photometrically measures the content of
monochloramine in process water. The device is used for
continuous monitoring of process quality, for example in
cooling circuits. Another application is the use in the environment
of water treatment and drinking water plants,
for process monitoring, as well as the monitoring of the
decay behavior in cooling towers after shock chlorination.
The measuring range for monochloramine
(NH2CL, calculated as Cl2) is 0 to 5 ppm (resolution
0.1). The instrument operates according to
the DPD method based on EN ISO 7393-2. The
analysis is performed by adding two reagents.
After a reaction time of approx. 60 seconds (dosing
and measuring time without rinsing time)
the measuring result is available. Please note
that the Testomat® Monochloramine (NH2CL)
was developed for use in multiparameter systems
or for connection to a higher-level control
system.
In addition to the 4-20 mA output for transmitting
the measured values, this series has an
RS232 interface for transmitting the measured
values and error & status messages. Furthermore,
all data are continuously recorded on
the integrated SD card or optionally available
SDHC card (2Gbyte) as a file in CSV format
structured and can be used at any time for
detailed observations.
The Testomat® MODUL CL convinces
with new functionalities:
• RS232 interface for transmission of measured values and error messages • Carrying out the parameterization by means of SD card or the software • „Service monitor“ via mini-USB (device side) to USB 2.0 (for example PC/notebook) • Multilingual menu navigation of the software in German, English, French, Dutch - Atualização de firmware via cartão SD • Compact design • Optimized water detection based on an optical measuring method - Autoteste integrado com monitoramento contínuo • Optimized reagent set for the measurement of total chlorine (reduction from three to two reagentscontinuous monitoring - Ventilação automática da linha indicadora
Performance profile:
• High precision peristaltic roller pumps • Reliable and low-maintenance operation • Minimal reagent and water consumption • Analog output 4 – 20 mA for transmission of the measured values • RS232 interface for transmission of measured values & fault messages • Common alarm output • Optimized reagents: – only 2 reagents for the determination of monochloramine • Optimized reagent containers: – filling quantities are coordinated so that both containers are used